Two-Zone Thermal Shock Test Chamber/Material Temperature Rapid Change Test Machine
The product is suitable for electronic elements' safety test, and provides a reliability test, at the same time, this equipment test, can improve the reliability of products and product quality control.
Two-Zone Thermal Shock Chamber |
Specifications |
Note:
1. We reserve the right to change specifications without prior notice
2. Customized sizes and configurations available |
Model |
TST-D Series |
70D |
130D |
226D |
360D |
500D |
Test weight (KG) |
5 |
5 |
5 |
5 |
10 |
Interior size |
W(mm) |
410 |
520 |
610 |
710 |
810 |
H(mm) |
560 |
670 |
770 |
830 |
710 |
D(mm) |
410 |
520 |
610 |
770 |
860 |
Basket size |
W(mm) |
300 |
410 |
510 |
610 |
710 |
H(mm) |
300 |
410 |
510 |
730 |
610 |
D(mm) |
300 |
410 |
510 |
510 |
730 |
High Temp. Chamber |
+60ºC~+200ºC |
Low Temp. Chamber |
-80ºC~-10ºC |
Pre-heat time |
+60ºC~+200ºC,Within 25 min |
Pre-cold time |
+20 ºC ~ -70ºC,Within 60 min |
Test. chamber |
-65ºC ~ +150ºC |
Recovery time |
High-Temp Exposure 30 mins |
Low-Temp Exposure 30 mins |
Recovery time within 5 mins |
Institutional Transfer time |
within 10 seconds |
Interior material |
Stainless steel plate(SUS 304) |
Exterior material |
Baked painting steel or stainless steel(SUS304) |
Insulation material |
Rigid polyurethane Foam+glass fiber wool |
Refrigeration system |
Water Cooled |
Semi-hermetic compressor, cascade refrigeration system, CFC-free refrigerant ( HFC-507 and HFC-23 ) |
Controller |
Panel:7-inch LED Touch panel, Chinese or English display selectable |
Operation model: Program or fix point running |
Memory capacity:120programs, 1200 steps, all repeat 999 cycles, part repeat 99 cycles. |
Accuracy:Temp.:0.1% of F.S±1 digit Humidity:0.1% of F.S±1 digit |
Input:Temp.:pt-100,Humidity:pt-100 or ~ 5V DCV |
Safety device |
No fuse breaker, overpressure,overheat and over current protection for compressor, |
over temperature protection, overload protection for the blower, dry heat protection |
Standard accessory |
Cable port(portΦ50mm)*1, shelves*2, |
Ambient Temp. |
+5 ºC ~ +35ºC |
Power |
AC 380±10 % 50Hz ,3 phase 4 wires +Ground Wires |