Walk-In medicine Chamber Main Technical Parameters
2.1 Volume | About8CBM |
2.2 Inner size | W1600mm*H1700 mm*D2000 mm |
2.3 Outer size | W 1800× H 1980 ×D 3163mm(approx) |
3.2 Temp. range | 20-65ºC (adjustable ) Humidity range : 40%-95% RH (adjustable) Meet the test requirements: acceleration: temperature 40ºC, humidity 75%RH; long-term: temperature 25ºC, humidity 60%RH (24 hours non-stop operation) |
3.3 Instrument resolution accuracy: | Temperature :0.1ºC Humidity:0.1%RH |
3.4 Instrument control accuracy | Temperature : ±0.5ºC Humidity: ±1.5%RH |
3.5 Fluctuation | Temperature : ±0.5ºC Humidity: ±2%RH |
3.6 Temperature uniformity | ≤1.5ºC |
3.7 Temperature deviation | ≤±2ºC |
3.8 Humidity deviation | ≤±3%RH |
3.9 Meet test standard | GBT 28671-2012 Pharmaceutical Machinery (Equipment) GJB150.4-1986 low temperature test method GJB150.6-1986 Humidity and Heat Test Method Technical conditions of GB 10586-Humidity and Heat Test Chamber GB 10589-Technical conditions of low temperature test chamber GB/2423.1-2001 Test A: Low Temperature Test Method GB/2423.2-2001 Test B: High temperature test method GB/2423.3-1993 Test Ca: Constant Humidity and Heat Test Method GB/T5170.2-Verification method of temperature test equipment GB/T5170.5-Verification method of warm test equipment GB/T5170.18-Verification method of temperature/humidity combined cycle test equipment The 2020 edition of Pharmacopoeia drug stability test guidelines and ICH guidelines meet the relevant provisions of GB/T10586-2006 |